Dilute Thrombin Time Panel




Test Mnemonic

DTT

CPT Codes

  • 85730 - QTY (1)
  • 85670 - QTY (1)

Includes

  • Dilute Thrombin Time
  • APTT Screen

Performing Laboratory

Cleveland Clinic Laboratories


Specimen Requirements

Volume Type Container Collect Temperature Transport Temperature Special Instructions
1 mLPlasmaSodium citrate (Lt. Blue) Centrifuge, aliquot and freeze ASAP.Collection tube must be filled to total fill volume. Inadequately filled tubes will be rejected. Non-testing sites: Centrifuge sample, aliquot plasma into a separate tube and label with Epic Beaker labels. Specimens should be frozen (-20C or colder).

Minimum Specimen Requirements

Volume Type Container Collect Temperature Transport Temperature Special Instructions
N/A    Collection tube must be filled to total fill volume. Inadequately filled tubes will be rejected.

Stability

Environmental Condition Description
Ambient4 Hours
Frozen14 days at -20 C; 6 months at -70 C
RefrigeratedUnacceptable

Days Performed

Sun - Sat

Turnaround Time

6 hours

Methodology

Name Description
Clotting Assay 

Clinical Info

The DTT is used for measuring and monitoring direct thrombin inhibitors (DTI). The expected therapeutic range for the appropriate DTI (dabigatran, bivalirudin and argatroban) will be provided based on the medication selected when placing the order. DTT is not affected by optical interferences (hemolysis, icteria, lipemia), high or low fibrinogen, coumadin, enoxaparin or direct Xa inhibitors (e.g., apixaban, rivaroxaban). However, unfractionated heparin is known to interfere with the DTT assay.