Nickel, Serum




Test Mnemonic

NICKEL

CPT Codes

  • 83885 - QTY (1)

LOINC ®

5702-6

Includes

  • Nickel, Serum

Performing Laboratory

ARUP


Specimen Requirements

Volume Type Container Collect Temperature Transport Temperature Special Instructions
2 mLSerumNo additive (Royal Blue) AmbientPatient should be encouraged to discontinue nutritional supplements, vitamins, minerals, and nonessential over-the-counter medications (upon the advice of their physician). Separate serum from cells within 2 hours of collection and aliquot into a trace metal-free transport tube (ARUP # 43116). Do not use utensils (i.e., syringes, needles, or pipettes) in the collection or transfer of the sample, pour directly into transport tube.

Minimum Specimen Requirements

Volume Type Container Collect Temperature Transport Temperature Special Instructions
0.5 mL     

Stability

Environmental Condition Description
AmbientAfter separation from cells: Indefinitely
RefrigeratedAfter separation from cells: Indefinitely
FrozenAfter separation from cells: Indefinitely

Days Performed

Sun - Sat

Turnaround Time

2 - 4 days

Methodology

Name Description
Inductively Coupled Plasma / Mass Spectrometry (ICP-MS) 

Reference Range

Nickel, Serum
Sex Age From Age To Type Range Range Unit
       Normal<=10.0ug/L

Special Info

Patient Prep: Diet, medication, and nutritional supplements may introduce interfering substances. Patients should be encouraged to discontinue nutritional supplements, vitamins, minerals, and nonessential over-the-counter medications (upon the advice of their physician). Do not use utensils (i.e., syringes, needles, or pipettes) in the collection or transfer of the sample, pour directly into transport tube. Serum that takes longer than two hours to aliquot or is collected or transported in containers other than specified will be rejected. This test is New York DOH approved.

Clinical Info

Serum nickel testing is intended to detect potentially toxic exposure. Elevated results may be due to skin- or collection-related contamination, including the use of a noncertified metal-free collection/transport tube. If contamination concerns exist due to elevated levels of serum nickel, confirmation with a second specimen collected in a certified metal-free tube is recommended.