Iodide




Test Mnemonic

BIODIN

CPT Codes

  • 83789 - QTY (1)
  • 83018 - QTY (1)

LOINC ®

2494-3

Aliases

  • Iodine

Performing Laboratory

Mayo Clinic Dpt of Lab Med & Pathology


Specimen Requirements

Volume Type Container Collect Temperature Transport Temperature Special Instructions
1 mLSerumNo additive (Navy Blue) RefrigeratedDo not use betadine during venipuncture. Patient should have no gadolinium- or iodine-containing contrast media administered in the previous 96 hours. Allow specimen to clot for 30 minutes then pour serum into metal-free, screw-capped, polypropylene vial.

Minimum Specimen Requirements

Volume Type Container Collect Temperature Transport Temperature Special Instructions
0.3 mL     

Stability

Environmental Condition Description
Ambient21 days
Refrigerated21 days
Frozen21 days

Days Performed

Tue, Fri

Turnaround Time

2 - 5 days

Methodology

Name Description
Inductively Coupled Plasma / Mass Spectrometry (ICP-MS) 

Reference Range

Iodide
Sex Age From Age To Type Range Range Unit
       40 - 92 ng/mL 

Special Info

Allow specimen to clot for 30 minutes then centrifuge. Carefully pour serum into metal-free, polypropylene vial, avoiding transfer of the cellular components of blood. Do not insert a pipet into the serum to accomplish transfer, and do not ream the specimen with a wooden stick to assist with serum transfer. High concentrations of gadolinium and iodine are known to interfere with most metals tests. If gadolinium-containing or iodine-containing contrast media have been administered, a specimen cannot be collected for 96 hours.

Clinical Info

This test is useful for determination of iodine overload and monitoring iodine levels in individuals taking iodine-containing drugs. Values between 80 ng/mL and 250 ng/mL have been reported to indicate hyperthyroidism. Values > 250 ng/mL may indicate iodine overload.